Hello all,



I have just finished repairing a Rohde & Schwarz URV5 Millivolt meter that had two problems. The first was that it was blowing fuses - this turned out to be a shorted 1000uF bulk capacitor in the power supply. But, after the repair it now gets a E051 on power up. Lacking any service or operating manual, I shot-gunned the unit by swapping the CPU board, then the few socketed chips located on the CPU board between a running unit I had, and this one. This allowed me to localize the problem to the 2kx8



52B13 EEPROM located in socket D8. I'm guessing that it might have become corrupted when the cap shorted. Anyway, I then read the contents of the good device and programmed the "bad" one successfully. Now it seems to power up OK, with no error codes. Can anyone with a service/operating manual tell me if the calibration data is stored in this device? I'd rather not have to send the unit out for calibration unless I need to.

Thanks!


Hi JW,

I've also an URV5 which need to be re-calibrated and I'm trying to get a copy of the service manual. May be we can share... ;-)

pf f5bqp

"JW" a écrit dans le message de news: snipped-for-privacy@4ax.com...

I have the same issue , I mean looking for a URV 5 user + service manual patrick - f6iib

formatting link

Hi ,

I'm still trying to get a copy for theservice manual for this rig! Thanks in advance for the help

pf F5BQP

"patrick" a écrit dans le message de news:

4789eb34$0$1218$ snipped-for-privacy@news.free.fr...

Pierre-Fran=E7ois wrote:

You are easily duped. In the future, I suggest you withhold your thanks until you actually get some.

In addition, please don't top-post in these groups; trim, then bottom-post or middle-post.

formatting link

Apparently you can order the required docs. (in english) via

formatting link

Greetings, Nap - Gent - Belgium

at

r
e

Added info:

If you are quick, you can react for the manual Service und C/ Kalibration:

formatting link
apparently in english and german

On the probes URV5-Z1, URV5-Z7, URV5-Z2 and URV5-Z4:

formatting link

7.pdf The second part is in english.

Greetings, Nap - Gent - Belgium

at

d

ter

en

I

ome

e
78&ssPage...

Also see:

formatting link
However, I am not sure how much of that one covers URV-5

Greetings, Nap - Gent - Belgium