Hello all,
I have just finished repairing a Rohde & Schwarz URV5 Millivolt meter that had two problems. The first was that it was blowing fuses - this turned out to be a shorted 1000uF bulk capacitor in the power supply. But, after the repair it now gets a E051 on power up. Lacking any service or operating manual, I shot-gunned the unit by swapping the CPU board, then the few socketed chips located on the CPU board between a running unit I had, and this one. This allowed me to localize the problem to the 2kx8
52B13 EEPROM located in socket D8. I'm guessing that it might have become corrupted when the cap shorted. Anyway, I then read the contents of the good device and programmed the "bad" one successfully. Now it seems to power up OK, with no error codes. Can anyone with a service/operating manual tell me if the calibration data is stored in this device? I'd rather not have to send the unit out for calibration unless I need to.Thanks!