Pimp my pH input

Hi,

I am trying to debug the first stage of a pH input circuit. The first stage is just a voltage follower made from a TL084

The pH probe centre electrode is connected directly to the non-inverting input of one opamp of a TL084 (this connection does not touch the PCB, it's all in mid air) The pH probe centre electrode also has a 33pF ceramic capacitor to ground, close to the TL084 The pH probe outer electrode is connected to ground There is a 1M feedback resistor from the output of the opamp to the inverting input.

When I connect a pH probe, the output of this first stage is about 75mV. According to the TL084 datasheet it has an input offset voltage of 6mV so I can't understand where the 75mV comes from . I have tried 2 different pH probes, one old one new, from different manufacturers, one with pointy tip and one with round bulb, but they both give the same result so I don't think it's the ph sensor.

The power supply I think is my problem, I decided to use the +10V and -10V from a MAX232 rather than use a chip like a ICL7660. The MAX232 can only supply the +V supply of the TL084 with 5.77V and the -V supply with -3.88V.

I am trying to calculate whether this is the problem. With this power supply, the 'ground' for the opamp is actually at 0.945V relative to circuit ground. With the pH probe in a pH 7 calibration solution, the output between electrodes should be close to 0V, so in this case from the opamp's point of view both inputs are at -0.945V, and the only signal on the output should be the offset voltage + (common mode voltage * CMRR).

The CMRR on the datasheet is 86dB, or 1 / 20,000. If the common mode voltage on the inputs is 0.945V the contribution to the output should only be 47uV i.e. nothing.

So, why is the first stage giving 75mV? Should I use the supply-voltage rejection ratio instead? It's also 86dB so adding them together still doesn't explain it.

The output of the first stage is connected via a 510K resistor to the next stage so it is hard to imagine that that is affecting it.

Any ideas would be much appreciated.

Reply to
Fox one! Fox one!
Loading thread data ...

Can you post the circuit somewhere and the specs on the probe?

Tom

Reply to
Tom Biasi

A pH probe set is a kind of electro-chemical cell, with one fairly low resistance connection to the solution (the reference probe that makes a connection to the solution with minimal voltage drop) and one very high resistance connection that includes the voltage offset produced by the cell. The hydrogen ion (naked protons that are much smaller than any atom) concentration of the solution determines the voltage produced by the cell, as they diffuse through the glass bulb of the measurement probe and produce a net voltage across the glass membrane.

You must pass nearly zero current through either of these connections to get an undistorted measure of the voltage difference produced by the protons. It sounds like you have a very high resistance load (very limited current through) the measurement probe, but how about the reference probe. If the solution has other ground paths, you may be passing current through it, and getting some voltage drop through it. Is the test solution in a glass container (and electrically isolated from any current path that includes the reference probe) or is it grounded through other means than just the reference probe?

Reply to
John Popelish

See for example

formatting link

Mr.pH --

formatting link
formatting link

Reply to
Mr.pH

Hi Tom,

Here is the circuit.

1M ___ .-----------|___|---------. | | | | | 5.77V | | | | | | | | | | | | pH probe | |\\| | .-----. '-----------|-\\ | | G | | >----------o------------- |_-_-_|--------------------|+/ | | | |/| TL084 '--o--' --- | | --- | | | | | 0V 0V |

-3.88V

(created by AACircuit v1.28.6 beta 04/19/05

formatting link

Reply to
Fox one! Fox one!

Are the DC supplies batteries or AC derived supplies? If the latter, you have to include the AC paths through the inter winding capacitances back to the line and from there to ground and from there, back to the zero volt node.

Have you checked to see if there is any AC hum on the opamp output?

Reply to
John Popelish

Thankyou that is a good page, I have read it before. My circuit is very similar except it is unity gain. The only reason I have the resistor in the feedback at all is that I read on this page:

formatting link
(search for "ph") that if the resistor has the same impedance as the probe, the voltage across it due to the opamp leakage current will match and thus cancel the voltage error due to the same leakage through the probe.

I forgot to mention to Tom the specs of the probe, but no I don't know what they are. I am trying to make it work with typical probes, which from what I've heard are round bulb types and have a impedance of about 400Mohms.

Reply to
Fox one! Fox one!

Fox one! Fox one! wrote: (snip)

The probe resistance is not known and very high, compared to

1 meg. You need an opamp that has a bias current so low that it produces an insignificant drop across the 100 meg or so probe resistance. Then you can eliminate the feedback resistor. I would go for an opamp with sub pico ampere bias current. perhaps the LMC6001 with 25 fempto amps of bias current.
formatting link
Reply to
John Popelish

My understanding of the probes is that the junction and the reference are all in the one package with only two terminals. In my circuit one terminal is connected to ground and the other directly to the non-inverting input of a TL084, so that should be zero current shouldn't it?

It does respond to ph changes, I should have mentioned this: In a pH 7 buffer the output is 75mV (expecting 0mV) In a pH 4 buffer the output is 195mV (expecting 180mV, as output should change roughly 60mV for each pH)

The test solutions are in disposable plastic cups with no other wires in them.

.
Reply to
Fox one! Fox one!

That is just comp[act [packaging. There are still two distance probes that contact the solution, one through a direct electrical connection through a fluid bridge, and one through a membrane.

I see. That simplifies things from a DC standpoint. Is the cup sitting on a metal sink or a non conductive surface? In a separate post, I have questions for you about the possible hum being injected through the supply.

Reply to
John Popelish

I've noticed that if I ground the ph input the opamp gives 1.7mV!! Maybe the probe is the problem?

Reply to
Fox one! Fox one!

Good point I just thought since the space is there I would put in the largest value I have in my parts pile. I tried bypassing it (I am beginning to get superstitious now) and it made no difference (of course I knew it wouldn't).

Ah yes I have seen that part but this is a project for a hobby community on the net, so I am trying to use parts you can get from the local electronics store in case they ever have to repair it. My local store stocks these which seem like they are the right type (FET input):

CA3130 5pA bias current CA3140 10pA TL084 30pA (what I am using) TL074 65pA

By my calculations, assuming the probe has an impedance of around 675Mohm (based on 'standard glass', hemispherical shape on this page:

formatting link
then a worst case

30pA leakage current would cause a voltage drop of 20mV. That is surprisingly large but still doesn't explain where 75mV is coming from.
Reply to
Fox one! Fox one!

I've tried a DC supply and an AC supply (my PCB supports both) but both give the same results. In normal operation I'd expect it to be used with an AC supply, and for it to be installed in an earthed metal case, with circuit ground connected to the case by the screw mounting points. i.e. circuit ground should aways be mains earth

I've probed it with my CRO and no I can't see any mains frequency on the output at all.

Thanks for your help.

Reply to
Fox one! Fox one!

They are sitting on top of my computer. I tried putting a book between them and lifting the cup upwards but that made no difference.

Reply to
Fox one! Fox one!

Then I am fresh out of leads to follow. Sorry. Perhaps your probe set is just contaminated or the reference probe plug clogged and is producing the 75 mV offset, no matter what amplifier is used to buffer it. I have seen probe sets go bad this way. You might try leaving the probe in the pH

7 buffer solution for a few days and see if the offset drifts in either direction. No need to keep the amplifier on all the time, though it shouldn't hurt.
Reply to
John Popelish

messagenews:0s7ei.16532$ snipped-for-privacy@news-server.bigpond.net.au...

Or your opamp has a lot higher bias current than it is supposed to have. Replace the probe with a 10 meg resistor and see how high the output goes.

Reply to
jpopelish

Sheesh that was the problem! I don't have a 10meg resistor but I replaced the TL084 and voila, it works. I could swear I already tried this, god knows I changed everything else... Thanks a lot for your help.

Reply to
Fox one! Fox one!

:-)

Live long and prosper.

Reply to
John Popelish

ElectronDepot website is not affiliated with any of the manufacturers or service providers discussed here. All logos and trade names are the property of their respective owners.