Tektronix 2467B (similar to 2465B): Error Message TEST 04 FAIL 13 (3453)

Dec 01, 2006 Last reply: 3 years ago 11 Replies

Hi,



i am owning an Tektronix 2467B. Maybe someone can give me some advice, please. The 2467B shows following message: TEST 04 FAIL 13 (3453) I have got the complete service manual, but it says something like "Option Error" ??? (The scope includes Option 10 and 11.)



If i delete the error message by pressing the A/B tigger button, the device works, but it has a awful DC-Offset on all channels (which varies with the V/div- selection).


Best regards from Berlin



Harald



"tek_user" wrote in news: snipped-for-privacy@79g2000cws.googlegroups.com:

Did you try running a DC balance? (IIRC,hold down both AC/DC/GND buttons until the DCBAL procedure starts)

IIRC,"test 04" failures involve NVRAM memory retention.(all your cal constants)

You may have to replace the NVRAM and recal the entire instrument.

Or you have bad preamp hybrids.

Jim Yanik jyanik at kua.net

Hi Jim,

thank you very much. After running the Dc-balance the offset is ok ! ...but the error message is still the same. I forgot to say, that after deleting the error message, there are a lot of question marks on the display: for example

50mV???5V????......

Is it possible to post a foto in this group ? I would be very glad, to get some advices. Sorry i was wrong it is an 2467 not 2467B.

Thanks

Jim Yanik schrieb:

"tek_user" wrote in news: snipped-for-privacy@16g2000cwy.googlegroups.com:

I'd say you have to perform a full recal,and if this occurs again,your NVRAM IC is bad,needing replacement and another recal.

Jim Yanik jyanik at kua.net

Hi,

i forgot to say that after runnig the DC-balance cal. the error message changed to :

TEST 04 FAIL 03 (3453)

The following things are still up to date:

. After running the Dc-balance the offset is ok !

You should ask this in sci.electronics.repair. There's a fellow who posts there who used to work for Tektronix. (sci.electronics.repair added)

JW schrieb:

Well, we also read here :)

Looks like a problem in the calibration constants, stored in the NVRAM A "TEST 04 FAIL 13 " condition means according to my manauls a "bad checksum in the NVRAM".

If the NVRAM is ok, than this error can be corrected by doing the calibration procedure (all steps mentioned in the handbook) and storing the constants in the last step into the NVRAM.

hth, greets from Stuttgart,

Andreas

Andreas Tekman schrieb:

Ah , sorry, noticed the follow up too late.

Andreas

Hei Andreas,

kannst Du mir bitte Deine E-Mail-Adresse mitteilen ? (Die hier im Forum hinterlegte funktioniert leider nicht.)

Gruss

Harald

Andreas Tekman schrieb:

tek_user schrieb:

Nicht leider, sondern das ist von mir so gew=FCnscht und beabsichtigt.

It is likely to be a run-down battery. The calibration constants are held in a CMOS RAM which needs battery voltage to hold the calibration constants.

Useful, but this goes back to 2007.

Join the Discussion

Have something to add? Share your thoughts — no account required.

Didn't find your answer?

Ask the community — no account required