tektronix 2465B test 04 fail 10

my scope show test 04 fail 10 J'm looking for schematic for this scope ( J have manual without diagram schematic ) probably scope need calibration or someting is damaged with memory

( sorry for my english ) Best regards from Poland

Reply to
rubens
Loading thread data ...

"rubens" wrote in news: snipped-for-privacy@localhost.talkaboutelectronicequipmen t.com:

Bad EAROM perhaps.That means you may have lost your calibration constants that the microprocessor loads into registers throughout the scope to set gains,timing speeds,etc.

--
Jim Yanik
jyanik
at
kua.net
Reply to
Jim Yanik

news: snipped-for-privacy@localhost.talkaboutelectronicequipmen

diagram

constants

set

04 fail 10 means bad checksum, bit 4 set. Or in other words: He has to run through the cal process.

The OP can download the service manual for free from

formatting link

This manual includes all the schematics and calibration process steps. Good luck!

hth, Andreas

Reply to
tekamn

Thank's for answers. I will try to calibrate scope . If it will not set then EAROM is bad. Where is possible it to buy EAROM?

Reply to
rubens

"tekamn" wrote in news: snipped-for-privacy@c13g2000cwb.googlegroups.com:

If he recals,and then gets another fail message,then the EAROM is bad;it's not retaining the data. We used to replace a lot of EAROMs at TEK while I was there.

--
Jim Yanik
jyanik
at
kua.net
Reply to
Jim Yanik

This manual (

formatting link
it's without the schematics :( Can these schemes has anyone?

Reply to
rubens