Does anyone still make/sell Weston (a.k.a. saturated cadmium - I think !) voltage reference cells & if so, who ? I've recently acquired a vintage thermocouple potentiometer & restoration depends on being able to get a replacement cell. Googling produces a lot of theory & history but no practical supplier solutions.
Weston Cells
Jun 14, 2006
22 Replies
I last bought one (for such a portable potentiometer) about 15 years ago. They are probably still floating around on the used market. They last a very long time if not broken or abused.
I can imagine some ways to make a modern replacement, certainly good enough for such a calibrator application, that wouldn't need a power switch, but it wouldn't be totally trivial. It would also need a reference for calibration.
Best regards, Spehro Pefhany
"it\'s the network..." "The Journey is the reward"
speff@interlog.com Info for manufacturers: http://www.trexon.com
Embedded software/hardware/analog Info for designers: http://www.speff.com
formatting link
includes enough information about the Weston standard cell to let me build one - if I had access to a chemical laboratory.
It includes a reference to a paper which may include more specific information.
[9] F. B. Silsbee, Establishment and Maintenance of the Elec-trical Units, NBS Circular 475, National Bureau of Standards, Washington, DC (1949).You need mercury, cadmium, cadmium sulphate. mercurous sulphate (all poisonous) and water (only life-threatening in large quantities). IIRR the standard cell was built up out of 15mm diameter soft soda-glass tubing to produce a more or less H-shaped cell - the tops of the H were open, and each of the bottoms was closed around a short length of platinum wire, which became your terminals.
One side of the H was loaded with enough mercury to cover the Pt wire, the other with amalgamated cadmium (cadmium metal dissolved in mercury.
You then dumped enough mercurous sulphate on top of the pure mercury electrode to cover the surface while staying below the crossbar of the H, and a similar amount of cadmium sulphate on top of the cadmium amalgam electrode.
Once you had done that you poured in enough saturated cadmium sulphate solution to fill the cell above the crossbar. Once the solution has percolated through the cadmium sulphate crystals (above the amalgam electrode) and the mercurous sulphate crystals (above the pure mercury electrode) you should have a working cell.
IIRR the open tops of the H cell were closed in some way to prevent evaporation and contamination, but I can't remember how.
Bill Sloman, Nijmegen
I think I saw one on sale a while back at:
L.F. Hanney
77, Lower Bristol Road, Bath BA2 3BG England44 (0) 1225 424811
~ Adrian Tuddenham ~
(Remove the ".invalid"s and add ".co.uk" to reply)
www.poppyrecords.co.uk
You can build one but I don't think you can buy them any more because the chemicals are just too hazardous. What you actually need is a reference voltage of 1.0183 V, right? Did you know that NIST and CERN now use Zener circuits (as a secondary standard calibrated to a Josephson junction) because the latter is more stable than a Weston cell? You can DIY but zener diodes have tolerance issues.
If I wanted a chemical primary voltage standard I suppose I'd start stacking coins and paper disks soaked in saltwater.
OTOH do you absolutely need that last .0183 of a volt? If not:
formatting link
Mark L. Fergerson
If *I* were wanting a voltage reference on the cheap I think I'd use one of the many bandgap chips that are available, add a heater so that ambient didn't enter into the picture, then calibrate it against a lab standard.
...Jim Thompson
| James E.Thompson, P.E. | mens |
| Analog Innovations, Inc. | et |
| Analog/Mixed-Signal ASIC\'s and Discrete Systems | manus |
| Phoenix, Arizona Voice:(480)460-2350 | |
| E-mail Address at Website Fax:(480)460-2142 | Brass Rat |
| http://www.analog-innovations.com | 1962 |
I love to cook with wine. Sometimes I even put it in the food.
...
If I were wanting to do it, I'd save myself a lot of time and buy a good one from a company just down the road from you, Jim:
formatting link
With initial accuracy under a millivolt on some models, I might not even worry about having it calibrated further. (I don't think the stacked coins and salt water would be worth the effort...I remember just a bit too much from my physical chemistry classes about the things that affect electrochemical potential. The Weston cell is built the way it is for some very specific reasons.) Thaler does sell to 'sperimenters in small (unit) quantities.
Cheers, Tom
The standard cell has a fairly high tempco compared to most decent modern references (something like 50ppm/K), so I'm not sure an oven is called for. Natsemi has some with a built-in heater and temp control, but they are fairly pricey ($60 or something like that). Even the Thaler ones are gross overkill.
But here's a source (hey, $50 off regular price!) and an incentive to design your own if you go with one that requires a power source:
formatting link
Maybe an 1.024V LM4140, with an op-amp buffer and trim network, but it probably will draw too much Iq to be left on all the time on batteries.
Best regards, Spehro Pefhany
"it\'s the network..." "The Journey is the reward"
speff@interlog.com Info for manufacturers: http://www.trexon.com
Embedded software/hardware/analog Info for designers: http://www.speff.com
Good link! They're probably using SiCr resistors and laser trimming. I didn't know they could be had that cheap! Probably a Burr-Brown spin-off... lots of spaciality companies in Tucson are.
...Jim Thompson
| James E.Thompson, P.E. | mens |
| Analog Innovations, Inc. | et |
| Analog/Mixed-Signal ASIC\'s and Discrete Systems | manus |
| Phoenix, Arizona Voice:(480)460-2350 | |
| E-mail Address at Website Fax:(480)460-2142 | Brass Rat |
| http://www.analog-innovations.com | 1962 |
I love to cook with wine. Sometimes I even put it in the food.
When I last looked - quite some time ago - the best buried zener references were a lot quieter and more stable than bandgap-based references.
Neither is going to be much use in a vintage thermocouple potentiometer, which won't have any kind of power supply, unless you sneak in a few lithium cells to power the voltage reference.
Bill Sloman, Nijmegen
Nah, just add a coil and a cow magnet, and shake it whenever you need to use it. ;-P
Cheers! Rich
Does not work, get error message:
*** Exception: Undefined REF. Parameters= in Filter (refid=0) ***
What about the use of those 5V regulators that Intersil makes (within
0.5mV)? Drive with a standard 9V battery, no switch and use a 510 ohm resistor in series with the output for short circuit protection (if feel it is necessary).
Don't you get buttermilk voltage then?
snipped-for-privacy@ieee.org wrote: ...
Really?? I would expect it to have some other power (typically a battery), because the thermocouple voltages will be well below the standard cell voltage, and you don't want to draw current from the standard cell. I would expect another battery driving the potentiometer through a rheostat. There would be a switch to set the galvanometer to compare the top of the potentiometer with the standard cell, and you'd adjust the rheostat to null that. Then switch the galvanometer over to compare the thermocouple voltage with the potentiometer output, adjust the potentiometer to null the galvo, and read the voltage. Double-check the ref to be sure. Read the included thermometer to correct the ref voltage... Ah, the "good old days." The aux power might even be arranged in a way that you really could use it to power the reference, too, though its voltage may be too low unless you have a reference that can run on a very low (e.g. 1.5V) input.
Cheers, Tom
High accuracy potentiometers did use separate power and voltage reference to avoid current draw from the voltage reference, but the typical thermocouple potentiometer used a mercury battery for both purposes (not an actual Weston cell, but similar). The battery was only connected when a momentary switch was held closed. Working Weston cells go for around $10 on eBay now and then, but they were not designed to handle the current draw of a potentiometer and probably wouldn't fit anyhow. Subsituting a modern battery and a voltage reference would be the best way to make this antique work; I would try it with an adjustable power supply first to see what voltage range the potentiometers calibration adjustment can accommodate, it might adjust to work with a standard voltage reference value like 1.0 V.
... Wow. Sounds like an inferior piece of junk. The only ones I ran across (in physical chemistry and physics labs) used standard cells only for calibration and avoided drawing any significant energy from them. The schematics I've found on the web are all for that type, and in fact show a resistor in series with the galvo, so you can keep the current low till you get close to balance, then a switch to short the resistor to get max sensitivity.
Why bother with a potentiometer with accuracy no better than you can get from a calibrated direct-reading D'Arsonval meter movement? Once you've gone to the effort to make a nice linear potentiometer, why not use it in a circuit that takes full advantage of it? Maybe marketing was as active "back then" as it is now. (Oh, yes. The theme of "The Music Man.")
Cheers, Tom
Yes, that's the kind I've run into. I did retrofit one with a regulator and some D alkaline cells to get rid of the unobtanium mercury D cells which powered the potentiometer. Current was only drawn (or put into) the standard cell during the calibration procedure- push a button down and null the galvo.
Best regards, Spehro Pefhany
"it\'s the network..." "The Journey is the reward"
speff@interlog.com Info for manufacturers: http://www.trexon.com
Embedded software/hardware/analog Info for designers: http://www.speff.com
Hi,
What about an old DVM?
I have dismantled a few over the years and come accross these strange merucry filled devices - look for an old bench top nixie tube base unit with
4 1/2 or 5 1/2 digits+. I seem to remember Solartron / Schlumberger were fond of these devices. Don't ever remember seeing one inside a Fluke or a HP though. These can be picked up for next to nothing at Ham Fest / Radio Rallies - you can even sell the Nixies on ebay and get your money back!Philip
They did actually have much better accuracy than any calibrated direct-reading D'Arsonval meter movement when used with thermocouples. While they draw current from the mercury battery, they do not draw any current from the thermocouple when nulled, that was the whole point. The mercury batteries were remarkably stable over a very long life, and their small tempco was compensated with a wire-wound thermistor in the voltage divider chain.
I worked in a lab which had a bunch of these potentiometers some decades ago, along with a hundred or so thermocouples all made from the same batch of wire, and the potentiometer/thermocouple combinations pretty much all came in within a few deg F prior to adjustment on their annual ice/tin calibration at 449.54 F. These potentiometers could not be callibrated with a standard cell since they only went up to 30 mV, which was typical with thermocouple potentiometers, and could only be read to the nearest 10 uV anyhow, entirely adequate for the application.
My antique Guildline potentiometer is another beast entirely, with a
2V range in steps of 1 uV or .2V range in steps of 0.1 uV. The whole thing including 2 saturated standard cells is in a 30.00 C oven, and it needs an external 20.00000 mA current source. The external current source needs to be adjustable but not calibrated, the first step in making a measurement is to balance it with one of the standard cells. This is probably the sort once found in chemistry and physics labs, now found in the shops of foolish antique collectors like myself :-).Join the Discussion