LeCroy 9450A Oscilloscope

Hi, I have a LeCroy 9450A Oscilloscope. It has problem about triggering mode. LeCroy 9450A Oscilloscope displayed "NO or SLOW TRIGGER". It works well most of times. But sometime it does not run triggered mode but it run auto mode(not triggered). I can not fix this problem. Please let me know how to fix this problem without calling maker service. I have not service manuals for LeCroy 9450A Oscilloscope. Please give me any advices .

Kishikawa.

Reply to
kishikawa
Loading thread data ...

It may be obvious but are you sure the trigger level is properly set ? You get this message on a perfectly working scope if the trigger level is simply too high or too low...

Cheers, Robert

"kishikawa" a écrit dans le message de news: FHr0e.294$ snipped-for-privacy@news-virt.s-kddi1.home.ne.jp...

Reply to
Robert Lacoste

Thank you, Robert.

The trigger level is maybe set properly. LeCroy 9450A Oscilloscope works most time perfectly.

I set trigger level to half level of Probe calibrator output. Probe is connected to Probe calibrator output.

Probe and trigger mode is set to DC coupling mode. Probe is connected to Ch1 and trigger source is Ch1.

I have used other Oscilloscopes(Tek465,hp1980,Iwatsu SS-series). I used them in trigger mode. But LeCroy 9450A Oscilloscope sometime can not use in triggered(normal) mode.

Please Advice me another point.

Thank you Kishikawa.

----- Original Message ----- From: "Robert Lacoste" Newsgroups: sci.electronics.equipment Sent: Thursday, March 24, 2005 4:45 PM Subject: Re: LeCroy 9450A Oscilloscope

Reply to
kishikawa

It is of course quite difficult to establish a diagnostic remotely, but given your explanations it *should* work... Double check that there are no other complex triggering modes selected... Good luck, Robert

"kishikawa" a ecrit dans le message de news: FyC0e.298$ snipped-for-privacy@news-virt.s-kddi1.home.ne.jp...

Reply to
Robert Lacoste

Thank you, Robert.

I checked complex triggering mode-----is not selected.

There are 4 Lecroy Oscilloscopes near my sights.

3 of 4 Lecroy Oscilloscopes are in trigger mode problem same as mine. Maybe LeCroy 9450A Oscilloscope or similar type has same problem.

Someone experienced same problem which already fixed. So,I want any advice to fix problem.

Thank you Kishikawa

de

de

Reply to
kishikawa

Thank you All.

I maybe fixed LeCroy 9450A Oscilloscope problems.

1.100MHz Xtal is poor ----> change to 100MHz Xtal OSC.

2.ECL power supply is -3.6V ----> correct to -5.0V. Variable resister for voltage adjustment in power supply unit is bad.

Thank you Kishikawa

Reply to
kishikawa

ElectronDepot website is not affiliated with any of the manufacturers or service providers discussed here. All logos and trade names are the property of their respective owners.