Hello, I have a HP 8116A function generator out of order - E42 failure code - In order to repair it I need all informations about this function generator ...( block diagrams - eletricals schematics ....) and a simple guide line to find the failure .. Thanks a lot .. Best regards
The first and most obvious place to look for the manual is from the manufacturer. Download the manual from Agilent's web site
formatting link
--
Dave M
MasonDG44 at comcast dot net (Just substitute the appropriate characters in the
address)
"In theory, there isn't any difference between theory and practice. In
practice, there is." - Yogi Berra
Thanks a lot ... but I have this doc .... (of course my first idea is to go to the www HP /Agilent ...) - I am looking for the diagrams and electricals schématics... I don't know where I can have these doc ...
Happy near year....
"Michael A. Terrell" a écrit dans le message de news: snipped-for-privacy@earthl>>
As you already know, the manual from Agilent does not include the schematics. According to my manual (which has the schematics), this error code is:
The output amplifier is unable to supply a negative offset. Possible faults: Output amplifier A1 Offset DAC A2: U24, U25, U26, U17 Offset latches A2: U21, U22, U23, U43
I've fixed a few of these with this code, and IIRC the problem was an easy one to solve - the -24VDC supply had a problem. If yours is bad, check the following:
2200uF capacitor C4
CR2 bridge rectifier
U4 voltage regulator
Picofuse located in the -24VDC feed.
I'm currently posting the complete service manual to alt.binaries.schematics.electronic, it should be there by the time you look. Good luck!
ElectronDepot website is not affiliated with any of the manufacturers or service providers discussed here.
All logos and trade names are the property of their respective owners.