Solartron 7150 Manual needed.

Oct 27, 2004 1 Replies

Looking for an E-copy of a Solartron 7150 manual OR a clue as to why the unit is only responding to the keypad in columns as opposed to as it should as a matrix. ie keypad is 3col x 4row matrix but only 3 functions available, all keys in one column produce same reult. All functions available via GPIB and all scales/ranges in good order.


Hi

The User's Manual does not say anything about such a state ... in other words, this is not "expected behaviour". If the 7150 is _not_ in local lockout state, every single key (mode/range settings etc) should react as you expect them to do.

Just to be sure: did you disconnect it from the GPIB bus? You did not accidently - via GPIB - send it into calibration mode (where it would no longer react to the front keys either)?

Cheers + HTH,

- Joerg

joerg dot hau at swissonline dot ch * Lausanne, Switzerland http://homepage.sunrise.ch/mysunrise/joerg.hau/ "All standard disclaimers apply". remove the obvious from my address to reply