I have the following problem on my 2465A. When I am connect ch1 and ch2 to the calibrator, ch2 displays a rectangular signal and ch1 displays something that becomes a triangular signal when I increase the frequency. I have put pictures on:
" snipped-for-privacy@algosyn.com" wrote in news: snipped-for-privacy@a75g2000cwd.googlegroups.com:
you can remove the CH.1 attenuator,take it apart carefully(a plastic frame retains the relays),and use an eraser to LIGHTLY clean the switch pads on the ceramic substrate.
*Don't remove the gold plate.***Don't crack the substrate.**
the attenuators are succeptable to outgassing from plastics coating the contact pads. The other part of the switches are the relays,I would not attempt to clean those contacts,maybe spray with a zero-residue/zero lube contact cleaner.
you have to remove the front plastic edge trim to access the screws to pull the front panel,to access the screws to remove the attenuator. there's a procedure in the service manual for atten removal.
New attenuators are no longer available;TEK does NOT support this product series anymore.
I will have a look this wee-end. I have not the schematic of the scope. Do you think I can exchange some parts of Ch1 and Ch2 to locate where is the problem ?