Hello everyone. This is my first time posting to this group. I have a 496P that I picked up on eBay and of course, it is not functioning. I have the four manuals for it, however the two service manuals do not match and both do not match the actual unit. This unit appears to be a newer design as the layout appears cleaner, but serial number is low (B020866).

Overall the unit appears in very good condition for the age. Hour meter is at about 2000 now. Most of the features appear to be working and the readings I get from it are very close. I ran the internal computer tests for fun and this passes without any problems.

It would appear that the outside PLL loop of the 1st LO does not function correctly, maybe. The sreen displays Error 58 anytime it tries to lock. Looking at the schematic in the service manual I have and tracing out the system, it does not appear that the FM coils negative side is connected. All the hardware is in place to drive the positive side of the coil. I assume that because both the service manual and the unit are the same that all of the tuning was with the main coil.

All of the frequencies appear correct in the innner loop. Looking at the 25MHz and the 5MHz strobe signal, they apppear to be locked and the IF on Q1040 of the offset mixer is at 50% and the frequency does appear to match with what I see on jumpers K and N.

Removing the 5MHz Stobe enable jumper (P3057) and setting the PLLs loop gain, I can make the adjustment to remove what I and the manual calls the beat note from (TP2038, on my card but called TP2037 in the service manual) to allow the Lock signal to set true. Of course there are no waveforms showing what the beat note would look like, so I am not 100% sure this is what I am seeing. I traced the Lock all the way to the bus and indeed, when the processor reads bus (U4025 74240 pin 1/19 low), pin 9 (lock on instrument bus side) goes low. Up and down also are low on pins 15 and 17.

So, it is not just Lock that the computer needs to see to make it happy and I know it is not locked by the drift.

I suspect that the problem may be in the phase detector. What I am seeing on its output is a 40mV p-p signal that corrisponds to when the Strobe is active, riding on a -4.0868. The DC voltage depends on the inputs to the phase detector and if I set the system to Max, I get a signal that would appear to represent the error. What I am not sure about is the 40mV. With the span set to 50KHz (or even 1MHz for that matter) there does not appear to be enough signal to make the F ERROR output change states. I can set the Error Count Breakpoint pot (R1061) to get something out with a 5MHz span, but no lower. Because of this, I never see the Valid Count (pin 8 of U4025) get set.

What really bothers me is that between the the two manuals and the actual unit I have, Tektronix made a lot of changes to the PLL circuit. I am guessing that this was not done for the fun of it and that they had problems.

If you have any ideas, shy of telling me to get a different one, I would like to hear them. Thanks in advance.

Reply to
lecroy7200
Loading thread data ...

First suggestion is to go ask the question at:

formatting link

Been a long time since I worked on one of these. Built several from dumpster parts many years ago. My job was fixing the bad boards. Troubleshooting is a lot easier with a set of extender boards. The things are packed in so tight you can't get a probe on 'em.

Tek made a bunch of improvements in the 49x series over the years. They changed multiple boards at once. You have to have every board matching the firmware version. It's not easy to tell what you need. If you don't know the history, you may not have a complete/compatible set of boards/firmware. Many of the changes were relatively simple, but you'd need schematics for both to do the mods.

IIRC, the upgrade that provides the high accuracy timebase and multi-digit frequency readout made things much more difficult to troubleshoot, cause the processor complains about every little thing. That's why I never updated mine. mike

--
Wanted, Serial cable for Dell Axim X5 PDA.
Return address is VALID but some sites block emails
with links.  Delete this sig when replying.
FS 500MHz Tek DSOscilloscope TDS540 Make Offer
Bunch of stuff For Sale and Wanted at the link below.
MAKE THE OBVIOUS CHANGES TO THE LINK
http://www.geocities.com/SiliconValley/Monitor/4710/
Reply to
mike

ElectronDepot website is not affiliated with any of the manufacturers or service providers discussed here. All logos and trade names are the property of their respective owners.