why spectrum analyser and not oscilloscope

May 05, 2005 1 Replies

Hi, Why do not the oscilloscope used for noise measurement, where we can use the FFT feature of the scope. Most of the time it is seen that expensive spectrum analysers are used. Any comments on it and discussion are most welcomed.



a écrit dans le message de news: snipped-for-privacy@g14g2000cwa.googlegroups.com...

Because even high end oscilloscope have a 8 bit ADC, providing a dynamix range less than 1/250 in voltage, or around 24dB(in voltage), or around

48dB(in power). The worst spectrum analyzers have 80dB of dynamic range...

Robert

Join the Discussion