DSA602 defaults

hello

does anybody know how to set this scope to defaults.I had minimized the display brightness and now i see nothing and i don't know how to change the screen to be bright.please help anybody

seba

Reply to
sebastian
Loading thread data ...

It looks like there are 3 batteries, remove / test/ replace as needed.

Below is how aworking scope is initalised: Press Utility Initialise settings in the Utility menu Initalise settings in the Verify pop-up menu

Reply to
carneyke

I cribbed this from the Tek website.

Q. My 11402 has locked up. I have tried powering it off, then on again with no success. How can I resolve problem? A.

When the instrument's power is cycled, the instrument tries to return to the same conditions as when it was powered down. Unfortunately, that can include conditions like you are experiencing. Returning the instrument to factory default conditions will usually correct these problems. NOTE: All settings and waveforms will be erased when you reset the instrument to its default settings. This procedure to do this is as follows:

Turn off the instrument. Hold down the Waveform and Trigger menu buttons, while turning the instrument on. When the menu lights start flashing, the buttons can be released. The instrument should display a message indicating that factory defaults are being restored and everything except the number of power-ups, the serial number and the time/date has been erased.

regards Andrew

Reply to
Andrew