+++++++++++++++++++++++++++++++++++++++++Date: Mon, 28 May 2012 11:12:54 -0400 =46rom: Monty Solomon Subject: Class 1 Recall: Nicolet, Software Malfunction and Short Circuit
CareFusion Cortical Stimulator Control Unit: Class 1 Recall - Software Malfunction and Short Circuit
Nicolet Cortical Stimulator Control Unit, Nicolet C64 Stimulus Switching Unit (SSU) Amplifier and NicoletOne Software with Cortical Stimulator License U.S. Food and Drug Administration [Posted 02/23/2012]
AUDIENCE: Neurology, Risk Managers
ISSUE: CareFusion is recalling Nicolet Cortical Stimulator Control Unit, Nicolet C64 Stimulus Switching Unit (SSU) Amplifier and NicoletOne Software with Cortical Stimulator License for two reasons: the device's software incorrectly indicates stimulation is delivered to a different electrode than the one selected and a short circuit may develop between the cortical stimulator control unit and the stimulus switching unit amplifier. Both of these issues may result in the surgeon resecting the wrong brain tissue. The surgeon may also fail to resect pathological tissue, potentially leading to continued pathologic processes and the need for re-operations. ...