Oscillator stops when probed.

I've an old Tektronix 184 with an 10 MHz OCXO hooked up to a nuvistor tetrode (valve). The crystal is connected between the cathode and the first grid. The cathode end of the crystal connects to earth through a variable 6 to 10 micro Henry inductor.

2nd grid should be at about 50V (I think) as there's a 75V zener between there and the 125V anode supply, but in fact it measures at around 70V DC on my DMM???

The anode (plate) is connected to a tank tuned to the 5th harmonic.

Without a 10x scope probe at the junction between the inductor, cathode and crystal, the system produces an output, so the oscillator is clearly running when I don't try to probe it.

If I put the scope probe very near the junction, it acts as a aerial, and picks up the signal at a very low level. The moment I touch the connection with the probe tip the oscillator stops (sometimes I see the correct output of about 70V peak to peak as shown on the schematics for a fraction of a second).

I suspect something is wrong with the circuit, but can't work out what it might be ...

Why is this likely to be happening?

--
Dave Partridge (remove xwy from replyto email address)
Reply to
David C. Partridge
Loading thread data ...

No divider probe? That's likely the problem.

An oscilliscope may have 1meg of resistance across it's input, but when you start connecting a cable to that input you won't be seeing 1meg impedance at the other end. The capacitance of the cable will load down the circuit under test. The higher in frequency, the more significant the cable will load down the circuit.

The divider probe ensures that the circuit under test will not see the load of the cable. It will see a relatively small load. This is exactly what the probe is for, the fact that it divides the input level by 10 is only a byproduct of this.

Michael

Reply to
Michael Black

Even a 10x probe can sometimes load a crystal oscillator sufficiently to kill it. Is there a particular reason you need to probe the circuit at the crystal?

Reply to
stickyfox

OH! I didn't realise that an 10 meg loading along with the capacitance of the probe could quite normally shut a crystal oscillator down.

I was using a x10 probe, so the load shouldn't have been much.

I was just trying to check everything out according to the service manual and waveforms throughout the circuit as this piece of kit is new to me (but very old!). A through check through seemed in order.

The service manual clearly shows a (drawn) scope shot of the output on the xtal circuit, so I expected to be able to probe it. Maybe a 100x probe might do the trick, but I don't have one of those.

Reply to
David C. Partridge

"David C. Partridge" bravely wrote to "All" (19 Oct 05 16:30:45) --- on the heady topic of "Oscillator stops when probed."

DCP> Reply-To: "David C. Partridge" DCP> Xref: core-easynews sci.electronics.repair:345698

DCP> I've an old Tektronix 184 with an 10 MHz OCXO hooked up to a nuvistor DCP> tetrode (valve). The crystal is connected between the cathode and DCP> the first grid. The cathode end of the crystal connects to earth DCP> through a variable 6 to 10 micro Henry inductor.

DCP> 2nd grid should be at about 50V (I think) as there's a 75V zener DCP> between there and the 125V anode supply, but in fact it measures at DCP> around 70V DC on my DMM???

DCP> The anode (plate) is connected to a tank tuned to the 5th harmonic.

DCP> Without a 10x scope probe at the junction between the inductor, DCP> cathode and crystal, the system produces an output, so the oscillator DCP> is clearly running when I don't try to probe it.

DCP> If I put the scope probe very near the junction, it acts as a aerial, DCP> and picks up the signal at a very low level. The moment I touch the DCP> connection with the probe tip the oscillator stops (sometimes I see DCP> the correct output of about 70V peak to peak as shown on the DCP> schematics for a fraction of a second).

DCP> I suspect something is wrong with the circuit, but can't work out what DCP> it might be ...

DCP> Why is this likely to be happening?

If this is with the direct probe, then the stray and cable capacitance is surely damping the oscillation. If it is stalling with the 10x probe then the tube gain might be marginal. If you know your probe's input capacitance then use a series cap 10 or 100 times smaller in value to increase the impedance further but you will need to set the gain correspondingly higher by 10 or 100. A voltage divider works wonders here. BTW, what kind of trouble do you suspect having?

A*s*i*m*o*v

... This message transmitted on 100% recycled photons.

Reply to
Asimov

Not all 10x probes are alike. I believe 10x probes exist with tip capacitances anywhere between about 7pF and 40pF, maybe higher. The low capacitance ones tend to be pricier, but it makes a siginificant difference once you get up into that frequency range.

A random 100x probe won't necessarly have lower C than a random 10x probe.

-

----------------------------------------------- Jim Adney snipped-for-privacy@vwtype3.org Madison, WI 53711 USA

-----------------------------------------------

Reply to
Jim Adney

What kind of trouble - well, the output waveform after the pickup transformer is quite distorted, so I though I'd check things closer to the source so to speak.

Yes, the gain on the tetrode could well be marginal, if it's the original (likely) it is now almost 40 years old.

Looks like I may need to get a job lot of 7587s (almost but not quite pure unobtanium).

DCP> Reply-To: "David C. Partridge" DCP> Xref: core-easynews sci.electronics.repair:345698

DCP> I've an old Tektronix 184 with an 10 MHz OCXO hooked up to a nuvistor DCP> tetrode (valve). The crystal is connected between the cathode and DCP> the first grid. The cathode end of the crystal connects to earth DCP> through a variable 6 to 10 micro Henry inductor.

DCP> 2nd grid should be at about 50V (I think) as there's a 75V zener DCP> between there and the 125V anode supply, but in fact it measures at DCP> around 70V DC on my DMM???

DCP> The anode (plate) is connected to a tank tuned to the 5th harmonic.

DCP> Without a 10x scope probe at the junction between the inductor, DCP> cathode and crystal, the system produces an output, so the oscillator DCP> is clearly running when I don't try to probe it.

DCP> If I put the scope probe very near the junction, it acts as a aerial, DCP> and picks up the signal at a very low level. The moment I touch the DCP> connection with the probe tip the oscillator stops (sometimes I see DCP> the correct output of about 70V peak to peak as shown on the DCP> schematics for a fraction of a second).

DCP> I suspect something is wrong with the circuit, but can't work out what DCP> it might be ...

DCP> Why is this likely to be happening?

If this is with the direct probe, then the stray and cable capacitance is surely damping the oscillation. If it is stalling with the 10x probe then the tube gain might be marginal. If you know your probe's input capacitance then use a series cap 10 or 100 times smaller in value to increase the impedance further but you will need to set the gain correspondingly higher by 10 or 100. A voltage divider works wonders here. BTW, what kind of trouble do you suspect having?

A*s*i*m*o*v

... This message transmitted on 100% recycled photons.

Reply to
David C. Partridge

ElectronDepot website is not affiliated with any of the manufacturers or service providers discussed here. All logos and trade names are the property of their respective owners.