Ar+ laser Tube

Jan 27, 2016 1 Replies

This is Thekrayat and a grad student at the University of Arkansas I am working with an old Argon ion laser, so Coherent company is not support this product an more.



The problem with this laser is that the Cathode current is decreased to 27 A then the laser is turning off by itself with " Fired starter 5x" fault massage. The technical support thinks that the laser tube is old and need to replace by a new one. I have checked the water level, interlock, and tube pressure, and every thing is ok. Also, I think there is no problem with the resistance of the four fuses in the power supply.



Is there any way to fix this laser without changing the laser tube? Thanks


ort this product an more.

7 A then the laser is turning off by itself with " Fired starter 5x" fault massage. The technical support thinks that the laser tube is old and need t o replace by a new one.

hing is ok. Also, I think there is no problem with the resistance of the fo ur fuses in the power supply.

Cue Sam Goldwasser, you might look here,

formatting link
And try your question at sci.optics

George H.