Reference
This section describes the menus and operating details associated with each front-panel menu button.
Reference
Acquire
Push the ACQUIRE button to set acquisition parameters.
Key Points
If you probe a noisy square wave signal that contains intermittent, narrow glitches, the waveform displayed will vary depending on the acquisition mode you choose.

Reference
The next two topics describe each of the types of acquisition modes and their differences.
Sample. Use Sample acquisition mode to acquire 2,500 points and display them at the SEC/DIV setting. Sample mode is the default mode.

• Sample points Sample mode acquires a single sample point in each interval.
The maximum sample rate is 1 GS/s. At 100 ns and faster settings, this sample rate does not acquire 2,500 points. In this case, a Digital Signal Processor interpolates points between the sampled points to make a full 2,500 point waveform record.
Peak Detect. Use Peak Detect acquisition mode to detect glitches as narrow as 10 ns and to limit the possibility of aliasing. This mode is effective when at 5 µs/div or slower.
Reference

- Sample points displayed Peak Detect mode displays highest and lowest acquired voltage in each interval.
NOTE. If you set the SEC/DIV to 2.5 \mus/div or faster, the acquisition mode changes to Sample because the sample rate is fast enough that you do not need to use Peak Detect. No message displays to tell you that the mode was changed from Peak Detect to Sample.
When there is enough waveform noise, a typical peak detect display shows large black areas. The TDS 200-series oscilloscopes displays this area with diagonal lines to improve display performance.
Typical peak detect display TDS 200 peak detect display
68 TDS 200-Series Digital Oscilloscope User Manual


Reference
Average. Use Average acquisition mode to reduce random or uncorrelated noise in the signal you want to display. Data is acquired in sample mode, then a number of waveforms are averaged together. Select the number of acquisitions (4, 16, 64, or 128) to average for the waveform.
Scan Mode Display. When the SEC/DIV control is set to 100 ms/div or slower and the trigger mode set to Auto, the instrument enters the scan acquisition mode. In this mode, the waveform display updates from left to right. There is no horizontal position or trigger control of waveforms during scan mode.
Stopping the Acquisition. While acquisition is running, the waveform display is live. Stopping the acquisition freezes the display. In either mode, the waveform display can be scaled or positioned with the vertical and horizontal controls.
TDS 200-Series Digital Oscilloscope User Manual 69
Autose
The Autose feature automatically adjusts the controls to produce a usable display of the input signal.
Pushing AUTOSET adjusts or sets each of the following items listed.
| Function | Setting |
|---|---|
| Acquire mode | Adjusted to Sample or Peak Detect |
| Vertical coupling | DC (if GND was selected) |
| Vertical VOLTS/DIV | Adjusted |
| Bandwidth | Full |
| Horizontal position | Centered |
| Horizontal SEC/DIV | Adjusted |
| Trigger type | Edge |
| Trigger source | Lowest numbered channel displayed |
| Trigger coupling | Adjusted to DC, Noise Reject, or HF Reject |
| Trigger slope | Rising |
| Trigger holdoff | Minimum |
| Trigger level | Set to 50% |
| Display format | YT |
| Trigger mode | Auto |
Reference
Cursors
Push the CURSOR button to display the measurement cursors and cursor menu.
- Only available on the TDS 224 oscilloscope.
Key Points
Cursor Movement. Use the CH 1 and CH 2 Vertical Position knobs to move cursors 1 and 2. You can move the cursors only while the Cursor menu is displayed.
Reference
Voltage cursors Time cursors
U in Level and Delta Readouts. Vertical sensitivity should match on waveforms used for math operations. If they do not match, and you use cursors to measure the waveform result of a math operation, an U displays that represents unknown.
Display
Push the DISPLAY button to choose how waveforms are presented and to change the appearance of the entire display.
Reference
Key Points
Persistence. When using persistence, old data retained is displayed in gray while the new data is black.
With Persistence set to Infinite, record points accumulate until a control is changed.
XY Format. This format only applies to channels 1 and 2. Choose XY display format to display channel 1 in the horizontal axis and channel 2 in the vertical axis. The oscilloscope uses the untriggered sample acquisition mode and data displays as dots. The sampling rate is 1 MS/s and can not be changed.
NOTE. The oscilloscope can capture a waveform in normal YT mode at any sampling rate. You can view the same waveform in XY mode. To do so, stop the acquisition and change the display format to XY.
The controls operate as follows:
■ The channel 1 VOLTS/DIV and vertical POSITION controls set the horizontal scale and position.
■ The channel 2 VOLTS/DIV and vertical POSITION controls continue to set vertical scale and position.
The following functions do not work in XY display format:
■ Ref or Math waveforms
■ Cursors
■ Autoset (resets display format to YT)
■ Time base controls
■ Trigger controls
Horizontal
You can use the horizontal controls to change the horizontal scale and position of waveforms. The horizontal center of the screen is the time reference for waveforms. Changing the horizontal scale causes the waveform to expand or contract around the screen center. Horizontal position changes the point, relative to the trigger, where the waveform appears on the center of the screen.
| Menu | Settings | Comments |
|---|---|---|
| Main | The main horizontal time base setting is used to display the waveform | |
| Window Zone | Two cursors define a window zoneAdjust the window zone with the Horizontal Position and SEC/DIV controls | |
| Window | Changes the display to show the waveform segment (expanded to screen width) within the window zone | |
| Trig knob | LevelHoldoff | Selects whether the Trigger Level knob adjusts the trigger level (volts) or holdoff time (sec)The holdoff value is displayed |
NOTE. You can push the Window menu box button to switch between an entire waveform display and an enlarged, more detailed part of it.
Reference
Key Points
SEC/DIV. If waveform acquisition is stopped (using the RUN/STOP button), the SEC/DIV control expands or compresses the waveform.
Scan Mode Display. When the SEC/DIV control is set to 100 ms/div or slower and the trigger mode is set to Auto, the instrument enters the scan acquisition mode. In this mode, the waveform display updates from left to right. There is no trigger or horizontal control of waveforms during scan mode.
Window Zone. Use the window zone to expand a segment of a waveform to see more detail. The Window time base setting cannot be set slower than the Main time base setting.
Vertical bars define window zone
Main time base displayed
Window zone displayed
Holdoff. Use holdoff to help stabilize the display of nonperiodic waveforms.
Holdoff begins when the instrument recognizes a trigger event and disables the trigger system until acquisition is complete. The trigger system remains disabled during the holdoff time that follows each recognized trigger event.

Reference
Math
Although you access the Math menu in the same way, the menu boxes vary, depending on the oscilloscope model, firmware version, and installed module.
Push the MATH MENU button to display the waveform math operations. Push the MATH MENU button again to turn off the math waveform display. Refer to page 89 for vertical system descriptions.
The following tables list menu boxes available based on the oscilloscope model, firmware version, and installed module.
TDS 210 and TDS 220 (Firmware V 2.00 and Above), and TDS 224 (All Versions) Oscilloscopes
- Only available on the TDS 224 oscilloscope.
NOTE. The optional TDS2MM measurement extension module adds the FFT feature to the oscilloscope. See the TDS 200–Series Extension Modules Instructions Manual.
Reference
TDS 210 and TDS 220 Oscilloscopes (Firmware Below V 2.00) without a TDS2MM
TDS 210 and TDS 220 Oscilloscopes (Firmware Below V 2.00) with a TDS2MM
The TDS2MM Module when used with a TDS 210 or TDS 220 changes the following math operations:
■ The TDS2MM Module does not have the CH1–CH2 and CH2–CH1 math functions. To do the same functions, invert the channel you want to subtract (using the CH1 or CH2 menu) and then select the CH1+CH2 math function.
Reference
| Default channel subtraction | TDS2MM channel subtraction | Comments |
|---|---|---|
| CH1-CH2 | CH1+(-CH2) | (-CH2) means invert channel 2 |
| CH2-CH1 | (-CH1)+CH2 | (-CH1) means invert channel 1 |
- The TDS2MM Module moves the channel invert function from the MATH menu to the Vertical CH1 and CH2 menus.
Key Points
VOLTS/DIV. Use the VOLTS/DIV control to scale the waveforms of the channels. The math waveform is the visual sum or difference of the channel waveforms.
Channel Display for a TDS 210 or TDS 220 (Firmware Below V 2.00). Displaying a math waveform automatically removes the display of channels used to create the math waveform. Math operations are turned off if a channel used in the operation is turned on.
NOTE. You can push the menu box button of the enabled math operation on a TDS 210 or TDS 220 to switch between the math waveform and the channels used to create the math waveform.
78 TDS 200-Series Digital Oscilloscope User Manual
Reference
Math Operations. Only one math operation is allowed. Using the subtraction selection eliminates the need to first invert and then add the waveform for subtraction operations.
Measure
Push the MEASURE button to access the automated measurement capabilities. There are five measurements available and the ability to display up to four at a time.
Press the top menu button to display the Source or the Type menu. You can choose the channel on which to take a measurement in the Source menu. You can choose the type of measurement to take (Freq, Period, Mean, Pk–Pk, Cyc RMS, and None) in the Type menu.

Source menu boxes Type menu boxes
NOTE. The optional TDS2MM measurement extension module adds the rise time, fall time, positive pulse width, and negative pulse width measurements to the oscilloscope. See the TDS 200–Series Extension Modules Instructions Manual.
TDS 200-Series Digital Oscilloscope User Manual 79
Reference
Key Points
Taking Measurements. You can display up to four automated measurements at a time for a single waveform (or divided among the waveforms). The waveform channel must be on (displayed) to make a measurement.
Automated measurements cannot be taken on reference or math waveforms or while using XY or scan mode.
| Measurement Type | Definition |
|---|---|
| Cyc RMS | Calculates a true RMS measurement of one complete cycle of the waveform |
| Mean | Calculates the arithmetic mean voltage over the entire record |
| Period | Calculates the time for one cycle |
| Pk-Pk | Calculates the absolute difference between the maximum and minimum peaks of the entire waveform |
| Freq | Calculates the frequency of the waveform by measuring the first cycle |
80 TDS 200-Series Digital Oscilloscope User Manual
Reference
Save/Recall
Push the SAVE/RECALL button to save or recall instrument setups or waveforms.
Setups
Key Points
Saving and Recalling Setups. The complete setup is stored in nonvolatile memory. When you recall the setup, you will be in the mode from which the setup was saved.
When turning the instrument on, all settings return to the settings that were in place when the instrument was turned off.
Recalling the Factory (Default) Setup. You can recall the default Factory Setup to initialize the instrument to a known setup.
Reference
Waveforms
- Only available on the TDS 224 oscilloscope.
Saving and Recalling Waveforms. For the TDS 210 and TDS 220 oscilloscopes, you can store two reference waveforms in nonvolatile memory. For the TDS 224 oscilloscope, you can store four reference waveforms, although you can only display two at a time. You can display reference waveforms simultaneously with current waveform acquisitions.
Recalled waveforms are not adjustable.
Reference
Trigger Controls
Two types of triggering are available: Edge and Video. A different set of menus display for each trigger type.
Edge Trigger
Use Edge triggering to trigger on the edge of the input signal at the trigger threshold.
- Only available on the TDS 224 oscilloscope.
** Only available on the TDS 210 and TDS 220 oscilloscopes.
Reference
Key Points
Normal and Auto Mode. Use Normal trigger mode to trigger only on a valid trigger. Use Auto trigger mode to let the acquisition free-run in the absence of a valid trigger. Auto allows an untriggered, scanning waveform at 100 ms/div or slower time base settings.
Single Mode. Use Single trigger mode to capture a single acquisition of an event. The content of a single acquisition sequence depends on the acquisition mode.
EXT and EXT/5 Source (TDS 210 and TDS 220 Only). The EXT trigger source uses the signal connected to the EXT TRIG connector. The EXT/5 trigger source uses the signal connected to the EXT TRIG connector and divides it by 5; this extends the trigger level range.
AC Line Source. The AC Line trigger source uses the power signal as the trigger source. Trigger coupling is set to DC and the trigger level to 0 volts.
Reference
Coupling. Coupling allows you to filter the trigger signal used to trigger an acquisition.
■ DC passes all components of the signal.
■ AC blocks DC components and attenuates signals below 10 Hz.
■ Noise Reject passes all components of the signal but increases the peak-to-peak signal required.
■ HF Reject attenuates the high-frequency components above 80 kHz.
■ LF Reject blocks the DC component and attenuates the low-frequency components below 300 kHz.
Pretrigger. The trigger position is typically set at the horizontal center of the screen. In this case you are able to view five divisions of pretrigger information. Adjusting the Horizontal Position of the waveform allows you to see more or less pretrigger information.
Trigger View. Pushing TRIGGER VIEW displays the trigger waveform until the button is released.
When in Trigger View mode, all front-panel buttons are disabled except HARDCOPY.
TDS 200-Series Digital Oscilloscope User Manual 85
Reference
Video Trigger
Choose video triggering to trigger on fields or the lines of an NTSC, PAL, or SECAM standard video signal.
| Menu | Settings | Comments |
|---|---|---|
| Video | With Video highlighted, triggering occurs on an NTSC, PAL, or SECAM standard video signalTrigger coupling is preset to AC | |
| Polarity | NormalInverted | Normal triggers on the negative edge of the sync pulse and Inverted triggers on the positive edge of the sync pulse |
| Source | CH1CH2CH3*CH4*EXT**EXT/5** | Selects the input source as the trigger signalEXT and EXT/5 use the signal applied to the EXT TRIG connector as the source |
| Sync | FieldLine | Select to trigger on fields or lines |
- Only available on the TDS 224 oscilloscope. ** Only available on the TDS 210 and TDS 220 oscilloscopes.
Key Points
Sync Pulses. When you choose Normal Polarity, the trigger always occurs on negative-going sync pulses. If your video signal has positive-going sync pulses, use the Inverted Polarity selection.
Utility
Push the UTILITY button to display the utility menus. The Utility menus change with the addition of extension modules. The menus explained here relate to the product with no modules installed. Refer to the manual supplied with your extension module for items not discussed here.
| Menu | Settings | Comments |
|---|---|---|
| System Status | Displays the system menus | |
| Do Self Cal | Performs a self calibration | |
| Error Log | Displays a list of any errors loggedThis list is useful when contacting a Tektronix Service Center for help | |
| Language | EnglishFrenchGermanItalianSpanishPortugueseJapaneseKoreanSimplified ChineseTraditional Chinese | Selects the display language of the operating system |
TDS 200-Series Digital Oscilloscope User Manual 87
Reference
Key Points
Self Calibration. The self calibration routine optimizes the oscilloscope accuracy for the ambient temperature. For maximum accuracy, perform a self cal if the ambient temperature changes by 5° C or more.
To compensate the signal path, disconnect any probes or cables from the input connectors. Then, select Do Self Cal to confirm that you are ready to proceed.
System Status
Selecting System Status from the utility menu displays the menus available for obtaining a list of control settings for each group of instrument controls.
Push any front-panel menu button to remove the status screen.
Reference
Vertical
You can use the vertical controls to display waveforms, adjust vertical scale and position, and set input parameters. Refer to page 76 for the vertical math descriptions.
Channel Vertical Menus
There is a separate vertical menu for each channel. Each item is set individually for each channel.
- Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
Reference
Key Points
GND Coupling. Use GND coupling to display a zero-volt waveform. When you use GND coupling, the input BNC connector is disconnected from internal circuits. Internally, the channel input is connected to a zero-volt reference level.
Fine Resolution. The vertical scale readout displays the actual Volts/Div setting while in the fine resolution setting. Changing the setting to coarse does not change the vertical scale until the VOLTS/DIV control is adjusted.
U in Level and Delta Readouts. Vertical sensitivity should match on waveforms used for math operations. If they do not match, and you use cursors to measure the waveform result of a math operation, an U displays that represents unknown.
Waveform off. To remove a waveform from the display, push the MENU button for the channel to display its vertical menu. Push the menu button again to turn the waveform off. An input channel can still be used as a trigger source or for math displays while turned off.
Hard Copy
Push the HARDCOPY button to print a hard copy of the display. The hard-copy function requires that an extension module with a Centronics, RS-232, or GPIB port be installed and connected to a printer.
Refer to the manual supplied with your extension module for instructions on connections and using the module.
Refer to Optional Accessories on page 103 for information about the available extension modules.