Appendix A: Specifications
All specifications apply to the TDS 200-Series Digital Oscilloscopes and a P2100 probe with the Attenuation switch set to 10X unless noted otherwise. To meet specifications, two conditions must first be met:
■ The instrument must have been operating continuously for twenty minutes within the specified operating temperature.
■ You must perform the Self Cal operation, accessible through the Utility menu, if the operating temperature changes by more than 5° C.
All specifications are guaranteed unless noted “typical.”
Specifications
Appendix A: Specifications
Specifications (Cont.)
| Input Impedance, DC Coupled, EXT TRIG only | TDS 210 (B099188 to B119999 and C021679 to C029999) and TDS 220 (B065810 to B079999 and C021127 to C029999) | All other TDS 210, TDS 220, and TDS 224 |
|---|---|---|
| 1.2 MΩ ±5% in parallel with 20 pF ±5 pF | 1 MΩ ±5% in parallel with 20 pF ±5 pF | |
| P2100 Probe Attenuation | 1X, 10X | |
| Probe Attenuation Factors | 1X, 10X, 100X, 1000X | |
| Maximum Voltage Between Signal and Common at input BNC | Overvoltage Category | Maximum Voltage |
| CAT I and CAT II | 300 VRMS (420 V peak, duty factor < 50%, pulse width < 100 msec.) | |
| CAT III | 150 VRMS | |
| For steady-state sinusoidal waveforms, derate at 20 dB/decade above 100 kHz to 13 Vpk at 3 MHz* and above. Also, refer to Overvoltage Category description on page 102. | ||
| Maximum Voltage Between Probe Tip and ground using P2100 probe connected to input BNC | Overvoltage Category | Maximum Voltage |
| CAT I and CAT II | 300 VRMS (500 V peak, duty factor < 35%, pulse width < 100 msec.) | |
| CAT III | 100 VRMS | |
| Derate at 20 dB/decade above 900 kHz to 13 VRMS at 27 MHz* and above. Also, refer to Overvoltage Category description on page 102. | ||
| Time delay between channels, typical | 150 ps |
- Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
92 TDS 200-Series Digital Oscilloscope User Manual
Appendix A: Specifications
Specifications (Cont.)
Inputs
| Channel-to-Channel Common Mode Rejection, typical | TDS 210 | TDS 220 and TDS 224 |
|---|---|---|
| 100:1 at 60 Hz20:1 at 30 MHz* | 100:1 at 60 Hz20:1 at 50 MHz* | |
| Measured on MATH Ch1 – Ch2 waveform, with test signal applied between signal and common of both channels, and with the same VOLTS/DIV and coupling settings on each channel. Also measured on MATH Ch3 – Ch4 waveform for the TDS 224. |
| Channel-to-Channel Crosstalk | TDS 210 | TDS 220 and TDS 224 |
|---|---|---|
| ≥ 100:1 at 30 MHz* | ≥ 100:1 at 50 MHz* | |
| Measured on one channel, with test signal applied between signal and common of the other channel, and with the same VOLTS/DIV and coupling settings on each channel. |
Vertical
| Digitizers | 8-bit resolution (except when set to 2 mV/div), each channel sampled simultaneously |
|---|---|
| VOLTS/DIV Range | 2 mV/div to 5 V/div at input BNC |
| Position Range | 2 mV/div to 200 mV/div, ±2 V> 200 mV/div to 5 V/div, ±50 V |
| Analog Bandwidth in Sample and Average modes at BNC or with P2100 probe, DC Coupled | TDS 210 | TDS 220 and TDS 224 |
|---|---|---|
| 60 MHz* (when vertical scale set to > 5 mV/div) | 100 MHz* (when vertical scale set to > 5 mV/div) | |
| 20 MHz* (when vertical scale set to ≤ 5 mV/div) |
| Analog Bandwidth in Peak Detect mode (5 s/div to 5 µs/div**), typical | TDS 210 | TDS 220 and TDS 224 |
|---|---|---|
| 50 MHz* (when vertical scale set to > 10 mV/div) | 75 MHz* (when vertical scale set to > 10 mV/div) | |
| 20 MHz* (when vertical scale set to ≤ 10 mV/div) |
- Bandwidth is not valid for the P2100 probe when the switch is set to 1X. ** The oscilloscope reverts to Sample mode when the sec/div (horizontal scale) is set from 2.5 µs/div to 5 ns/div. The Sample mode can still capture 10 ns glitches.
Appendix A: Specifications
Specifications (Cont.)
Vertical
- Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
** The oscilloscope reverts to Sample mode when the sec/div (horizontal scale) is set from 2.5 µs/div to 5 ns/div. The Sample mode can still capture 10 ns glitches.
Appendix A: Specifications
Specifications (Cont.)
Vertical
| Delta Volts Measurement Accuracy, Average Acquisition Mode | Delta volts between any two averages of ≥ 16 waveforms acquired under same setup and ambient conditions | ±(3% × reading + 0.05 div) |
|---|---|---|
| P2100 passive probe | 10X position | 1X position |
| Bandwidth | DC to 100 MHz | DC to 7 MHz |
| Attenuation ratio | 10:1 | 1:1 |
| Compensation Range | 18 pf-35 pf | All oscilloscopes with 1 M Ω input |
| Input resistance | 10 M Ω ± 2% | 1 M Ω ± 2% |
| Input capacitance | 14.5 pf-17.5 pf | 80 pf-110 pf |
| Maximum input voltage | 10X position | 300 VRMS CAT I or 300 V DC CAT I300 VRMS CAT II or 300 V DC CAT II100 VRMS CAT III or 100 V DC CAT III |
| 1X position | 150 VRMS CAT I or 150 V DC CAT I150 VRMS CAT II or 150 V DC CAT II100 VRMS CAT III or 100 V DC CAT III |
Appendix A: Specifications
Horizontal
Appendix A: Specifications
Specifications (Cont.)
Trigger
\begin{tabular}{|p{0.25\textwidth}|l|p{0.4\textwidth}|} \hline \textbf{Trigger Sensitivity, Edge Trigger Type} & \textbf{Coupling} & \textbf{Sensitivity} \ \hline & DC & CH 1, CH 2, CH 3 & CH 4 \newline 1 div from DC to 10 MHz*, 1.5 div from 10 MHz* to Full \ \cline{3-3} & & TDS 210 and TDS 220 only \ \cline{2-3} & EXT & 100 mV from DC to 10 MHz*, 150 mV from 10 MHz* to Full \ \cline{2-3} & EXT/5 & 500 mV from DC to 10 MHz*, 750 mV from 10 MHz* to Full \ \hline \textbf{Trigger Sensitivity, Edge Trigger Type, typical} & Coupling & Sensitivity \ \hline & AC & Same as DC at 50 Hz and above \ \cline{2-3} & NOISE REJ & Reduces the DC-coupled trigger sensitivity by 2 times for > 10 mv/div to 5 V/div \ \cline{2-3} & HF REJ & Same as the DC-coupled limit from DC to 7 kHz, attenuates signals above 80 kHz \ \cline{2-3} & LF REJ & Same as the DC-coupled limits for frequencies above 300 kHz, attenuates signals below 300 kHz \ \hline \textbf{Trigger Level Range} & Source & Range \ \hline & Internal & ±8 divisions from center of screen \ \cline{2-3} & & TDS 210 and TDS 220 only \ \cline{2-3} & EXT & ±1.6 V \ \cline{2-3} & EXT/5 & ±8 V \ \hline \end{tabular}
- Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
Appendix A: Specifications
Specifications (Cont.)
Appendix A: Specifications
General Specifications
Appendix A: Specifications
General Specifications (Cont.)
Appendix A: Specifications
General Specifications (Cont.)
Certifications and Compliances
| EC Declaration of Conformity | Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility*; meets the requirements of FCC Code of Federal Regulations, 47 CFR, Part 15, Subpart B, Class A: |
|---|---|
| (TDS 210, TDS 220, TDS 224) | CISPR 11 (Class A): Electromagnetic radiated and conducted emissions |
| EN 50082-1 European Community Requirements: | |
| IEC 61000-4-2 electrostatic discharge | |
| IEC 61000-4-3 radiated susceptibility** | |
| IEC 61000-4-4 power line fast transients | |
| IEC 61000-4-5 power line surge immunity | |
| IEC 61000-4-6 conducted immunity to RF fields | |
| IEC 61000-4-11 voltage dips, interrupts and variations | |
| AS/NZS 2064, Australian emissions standard for Industrial, Scientific, and Medical Equipment | |
| (TDS 210, TDS 220, TDS 224, P2100) | Low Voltage Directive 73/23/EEC as amended by 93/68/EEC: |
| EN 61010-1/A2:1995 Safety requirements for electrical equipment for measurement, control, and laboratory use | |
| (P2100) | EN 61010-2-031:1994 Particular requirements for hand-held probe assemblies for electrical measurement and test |
- Instrument tested with shielded cables.
** Meets Performance Criterion A with allowed degradation of up to 5 major divisions of trace noise increase from 80 MHz to 200 MHz and up to 2 divisions of trace noise increase over the range of 200 MHz to 1000 MHz under a 3 V/m RF field. Ambient RF fields may induce triggering when trigger threshold is offset less than 2.5 major divisions from ground reference.
Appendix A: Specifications
General Specifications (Cont.)
Certifications and Compliances
| Certifications (TDS 210, TDS 220, TDS 224, P2100) | CAN/CSA C22.2 No. 1010.1-92 UL3111-1, First Edition |
|---|---|
| (P2100) | CAN/CSA C22.2 No. 1010.2.031 IEC 61010-2-031 |
| CSA Certified Power Cords | CSA Certification includes the products and power cords appropriate for use in the North America power network. All other power cords supplied are approved for the country of use. |
| Pollution Degree 2 | Do not operate in environments where conductive pollutants may be present. |
| Overvoltage Category | Category: Examples of Products in this Category: CAT III Distribution-level mains, fixed installation CAT II Local-level mains, appliances, portable equipment CAT I Signal levels in special equipment or parts of equipment, telecommunications, electronics |
Adjustment Interval
The recommended calibration interval is one year
Appendix B: Accessories
All accessories (standard and optional) are available by contacting your local Tektronix field office.
Standard Accessories
P2100 1X, 10X Passive Probes. The P2100 passive probes have a 7 MHz bandwidth with a rating of 150 V CAT II when the switch is in the 1X position, and a100 MHz bandwidth with a rating of 300 V CAT II when the switch is in the 10X position.
TDS 200-Series Oscilloscope User Manual. A single User Manual is included. Refer to the Optional Accessories for a complete list of available language manuals.
Optional Accessories
TDS2MM Measurement Extension Module. The TDS2MM module plugs directly into the rear panel of any TDS 200-Series oscilloscope. This module provides full GPIB and RS-232 compatibility and a Centronics port for hard copy output. The module allows measurement of rise time, fall time, and positive and negative pulse width. The module also provides Fast Fourier Transform (FFT).
TDS2CM Communications Extension Module. The TDS2CM Communications module plugs directly into the rear panel of any TDS 200-Series oscilloscope. This module provides full GPIB and RS-232 compatibility and a Centronics port for hard copy output.




Appendix B: Accessories
Optional Accessories (Cont.)
RM200 Rackmount Kit. The RM200 Rackmount Kit lets you install a TDS 200-Series oscilloscope into an industry-standard 19 inch rack. The rackmount kit requires seven inches of vertical rack space. You can turn the oscilloscope power on or off from the front of the rackmount kit. The rackmount kit does not have slide-out capability.
TDS 200-Series Digital Oscilloscope Service Manual. The service manual (071-0492-XX, English) provides module level repair information.
TDS 200-Series Digital Oscilloscope User Manuals. The User manual is available in these languages:
English 071-0398-XX French 071-0400-XX* Italian 071-0401-XX* German 071-0402-XX* Spanish 071-0399-XX* Japanese 071-0405-XX* Portuguese 071-0403-XX* Simplified Chinese 071-0406-XX* Traditional Chinese 071-0407-XX* Korean 071-0408-XX* Russian 071-0404-XX
TDS 210 & TDS 220
TDS 224
*These manuals contain a language overlay for the front-panel controls.
104 TDS 200-Series Digital Oscilloscope User Manual
Appendix B: Accessories
Optional Accessories (Cont.)
International Power Cords. In addition to the power cord shipped with your instrument, you can obtain the following cords:
Option A0, North American 120 V, 60 Hz 161-0230-01 Option A1, European 230 V, 50 Hz 161-0104-06 Option A2, United Kingdom 230 V, 50 Hz 161-0104-07 Option A3, Australian 240 V, 50 Hz 161-0104-05 Option A4, North American 240 V, 60 Hz 161-0104-08 Option A5, Switzerland 230 V, 50 Hz 161-0167-00 Option AC, China 220 V, 50 Hz 161-0304-00
Soft Case. The soft case (AC220) protects the instrument from damage and provides space for probes, power cord, and manuals.
Transit Case. The transit case (HCTDS32) provides shock, vibration, impact and moisture protection for the instrument when you transport it from one place to another. The required soft case fits inside the transit case.


Appendix B: Accessories
106 TDS 200-Series Digital Oscilloscope User Manual
Appendix C: General Care and Cleaning
General Care
Do not store or leave the instrument where the LCD display will be exposed to direct sunlight for long periods of time.
CAUTION. To avoid damage to the instrument or probes, do not expose them to sprays, liquids, or solvents.
Cleaning
Inspect the instrument and probes as often as operating conditions require. To clean the exterior surface, perform the following steps:
- Remove loose dust on the outside of the instrument and probes with a lint-free cloth. Use care to avoid scratching the clear plastic display filter.
- Use a soft cloth dampened with water to clean the instrument. Use an aqueous solution of 75% isopropyl alcohol for more efficient cleaning.
CAUTION. To avoid damage to the surface of the instrument or probes, do not use any abrasive or chemical cleaning agents.
General Care and Cleaning
108 TDS 200-Series Digital Oscilloscope User Manual
Glossary
AC Coupling A mode that blocks the DC component of a signal but passes the dynamic (AC) component of the signal. Useful for observing an AC signal that is normally riding on a DC signal.
Acquisition The process of sampling signals from input channels, digitizing the samples, processing the results into data points, and assembling the data points into a waveform record. The waveform record is stored in memory.
Aliasing A false representation of a signal due to insufficient sampling of high frequencies or fast transitions. A condition that occurs when an oscilloscope digitizes at an effective sampling rate that is too slow to reproduce the input signal. The waveform displayed on the oscilloscope may have a lower frequency than the actual input signal.
Attenuation The degree the amplitude of a signal is reduced when it passes through an attenuating device such as a probe or attenuator (the ratio of the input measure to the output measure). For example, a 10X probe attenuates, or reduces, the input voltage of a signal by a factor of 10.
Auto Trigger Mode A trigger mode that causes the oscilloscope to automatically acquire if it does not detect a trigger event.
Average Acquisition Mode A mode in which the oscilloscope acquires and displays a waveform that is the averaged result of several acquisitions. This reduces the apparent noise. The oscilloscope acquires data as in the sample mode and then averages it according to a specified number of averages.
Glossary
Backlight The illumination behind the liquid-crystal display.
Bezel Buttons The column of buttons beside the display that select items in the menus.
Centronics A serial communication port used to connect to a hard-copy device, computer, controller, or terminal.
Cursors Paired markers that you can use to make measurements between two waveform locations. The oscilloscope displays the values (expressed in volts or time) of the position of each cursor and the distance between the two cursors.
DC Coupling A mode that passes both AC and DC signal components to the circuit. Available for both the trigger system and the vertical system.
Digitizing The process of converting a continuous analog signal such as a waveform to a set of discrete numbers representing the amplitude of the signal at specific points in time.
Display The word used to refer to the screen or the LCD (liquid crystal display).
Edge Trigger Triggering that occurs when the oscilloscope detects the source passing through a specified voltage level in a specified direction (the trigger slope).
GPIB A serial communication port used to connect to a hard-copy device, computer, controller, or terminal.
Glossary
Ground (GND) Coupling Coupling option that disconnects the input signal from the vertical system.
Ground Lead The reference lead for an oscilloscope.
Hard Copy An electronic copy of the display in a format useable by a printer.
Holdoff A specified amount of time that must elapse after a trigger event before the trigger circuit will accept another trigger event Holdoff helps ensure a stable display.
Voltage Cursors The two horizontal bars that you position to measure the voltage parameters of a waveform. The oscilloscope displays the value of each cursor with respect to ground and the voltage value between the bars.
Menu A set of labels shown in the display to identify the functions of the bezel buttons. The specific menu contents depend on the menu button you press.
Normal Trigger Mode A mode where the oscilloscope does not acquire a waveform record unless a valid trigger event occurs. It waits for a valid trigger event before acquiring waveform data.
Peak Detect Mode A mode in which the oscilloscope creates a pair of record points during each sample interval. Each pair consists of the maximum and minimum input voltage during the interval.
Persistence A method of retaining old data on the display for a period of time.
Glossary
Pixel A visible point on the display. The display is 320 pixels wide by 240 pixels high.
Pretrigger The portion of the waveform record that contains data acquired before the trigger event.
Real-Time Digitizing A digitizing technique that samples the input signal with a sample frequency of at least four times the oscilloscope bandwidth. (TDS 200-series oscilloscopes sample at 10 times the bandwidth or more.) Combined with (sinx)/x interpolation, all frequency components of the input up to the bandwidth are accurately displayed.
Record Length The number of samples in a waveform.
Reference Waveform A saved waveform selected for display.
RS-232 A serial communication port used to connect to a hard-copy device, computer, controller, or terminal.
Sample Acquisition Mode A mode in which the oscilloscope creates a record point by saving a sample during each acquisition interval.
Sample Interval The time interval between successive samples in a waveform record. Changing the SEC/DIV control (the time base) changes the sample interval. For real-time digitizers, the sample interval is the reciprocal of the sample rate.
Sampling The process of capturing an analog input, such as a voltage, at a discrete point in time and holding it constant so that it can be quantized.
Glossary
Scan Mode
An acquisition mode useful at slow horizontal scale settings. Scan mode allows you to view the waveform as it is acquired point-by-point. The waveform appears from left to right across the display.
Time Base
The period of time that corresponds to one horizontal division on the screen (one tenth of the waveform record) set with the SEC/DIV control. See also Sample Interval.
Time Cursors
The two vertical bars you position to measure the time parameter of a waveform record. The oscilloscope displays the value of each cursor with respect to trigger and the time value between the bars, expressed in seconds and in the reciprocal of seconds (Hertz).
Video Trigger
Triggering on the sync pulse of a composite video signal.
XY Format
A display format that compares the voltage level of two waveform records point by point. It is useful for studying phase relationships between two waveforms.
YT Format
The conventional oscilloscope display format. It shows the voltage of a waveform record (on the vertical axis) as it varies over time (on the horizontal axis).
Glossary
114 TDS 200-Series Digital Oscilloscope User Manual
Index
A AC line, trigger source, 11, 84 accessories, 103–106 acquire, menu, 66 ACQUIRE button, 34, 66 acquiring data, basic concepts, 15 acquisition live display, 69 modes, 15, 66 Average, 16 Peak Detect, 16 Sample, 15 single shot, 49 stopping, 69 aliasing, 18 applications analyzing a differential communication signal, 60 analyzing signal detail, 47 automatic measurements, 38 autoset, using, 38 averaging, using, 48 calculating amplifier gain, 41 capturing a single-shot signal, 49 cursors, using, 42 looking at a noisy signal, 47 measuring propagation delay, 51 measuring pulse width, 42 measuring ring amplitude, 46 measuring ring frequency, 45 measuring rise time, 43 measuring two signals, 40 optimizing the acquisition, 50 peak detect, using, 47 reducing noise, 48 taking automatic measurements, 39 taking cursor measurements, 42 triggering on a video signal, 53 triggering on odd or even video fields, 57 triggering on video fields, 54 triggering on video lines, 54 using persistence, 64 using the math functions, 61 using the window function, 55 using xy mode, 64 viewing impedance changes in a network, 63 attenuation, probe, 89 auto trigger, 12, 84 automated measurements, 21 AUTOSET, 22 actions, 70 button, 34, 70 average, 16, 66, 69
B bandwidth, limit, 89 basic concepts, 9 broken waveforms, 28 buttons, trigger view, 33
C calibration, 87 Centronics, 2, 103 CH 1 connector, 35
Index
MENU button, 30 CH 2 connector, 35 MENU button, 30 CH 3 connector, 35 MENU button, 30 CH 4 connector, 35 MENU button, 30 channel, scale, 26 cleaning, 107 coarse resolution, 89 compensation probe, 6, 35 signal path, 88 connectors, 35 CH 1, 35 CH 2, 35 CH 3, 35 CH 4, 35 EXT TRIG, 35 PROBE COMP, 35 contrast, 72 coupling trigger, 14, 83, 85 vertical, 89, 90 cursor adjusting, 34 menu, 71 using, 71 CURSOR button, 34, 71 cursors, 21 application example, 42 measurements, 42 time, 21 voltage, 21
D description, general, 1
display, 24 contrast, 72 format, 72 indicators, 24 intensity, 72 menu, 72 persistence, 72 type, 72 DISPLAY button, 34, 72 dots, 72 dual time base, 32, 74
E error log, 87 EXT, trigger source, 11 EXT TRIG connector, 35 EXT/5, trigger source, 11 extension, module, 2, 103 external trigger, 35
F factory setup, recalling, 81 features, overview, 1 fine resolution, 89 FORCE TRIGGER button, 33 format, 72 functional check, 5
G GPIB, 2, 103 graticule, 20, 72 gray waveforms, 28 ground marker, 26 probe terminal, 7 terminals, 35
116 TDS 200-Series Digital Oscilloscope User Manual
Index
H
HARDCOPY button, 34, 90 holdoff, 13, 32, 75 HOLDOFF control, 13, 32 horizontal aliasing, 18 menu, 74 position, 17 position marker, 25 scale, 17 status, 88 HORIZONTAL MENU button, 31
I
intensity, 72 invert, 77
L
languages, 87 LEVEL, control, 32 level, 15
M
main time base, 32, 74 manuals, 104 math functions, 76 menu, 76 MATH MENU button, 30 measure, menu, 79 MEASURE button, 34 measurements automatic, 21, 79 basic concepts, 20 cursor, 21, 42 frequency, 80 graticule, 20 mean, 80 peak-to-peak, 80 period, 80 RMS, 80 types, 80 menu boxes Action Button, 27 Circular List, 26 Page Selection, 27 Radio Button, 27 menu system menu boxes, 26 using, 26 menus acquire, 66 cursor, 71 display, 72 horizontal, 74 math, 76 measure, 79 save/recall, 81 trigger, 83 utility, 87 vertical, 89 messages, 26
N
normal operation, recall factory setup (default), 22 normal trigger, 84 NTSC, 57, 86
P
PAL, 54, 86 peak detect, 66, 67 persistence, 72, 73 ports, 2
Index
position, vertical, 89 POSITION control horizontal, 31 vertical, 30 positioning waveforms, basic concepts, 16 power, 3 power cords, 3, 105 pretrigger view, 85 printing, 90 PROBE COMP connector, 35 probes attenuation, 89 compensation, 6, 35 safety, 7 specifications, 101
R recalling factory setup (default), 22 setups, 22 waveforms, 82 resolution, fine, 90 roll mode. See scan mode RS-232, 2, 103 hard copy, 90 RUN/STOP button, 34
S sample, 66, 67 sample rate, maximum, 67 SAVE/RECALL button, 34, 81 saving setups, 22, 81 waveforms, 82 scaling waveforms, basic concepts, 16 scan mode, 69, 75 scanning waveform, 69, 75, 84 SEC/DIV control, 32, 75 SECAM, 86 security, 3 self cal, 87 service, 87, 104 SET LEVEL TO 50% button, 33 setups basic concepts, 21 saving and recalling, 81 signal path compensation, 88 single acquisition, 84 single shot, application example, 49 slope, 14 source, trigger, 10, 83, 86 specifications, 91 status, system, 87 sync pulse, 86
T time base, 26 main, 32, 74 window, 32, 74 time cursors, 21, 71 trigger coupling, 14, 83, 85 edge, 11, 83 external, 86 field, 86 holdoff, 13, 32, 75 level, 15, 32, 75, 83 marker, 25 readout, 25 menu, 83 modes, 11 Auto, 12, 84 Normal, 12, 84 Single, 12, 84
118 TDS 200-Series Digital Oscilloscope User Manual
Index
position, 14 position marker, 25 slope, 14, 83 source, 10, 83 status, 25, 88 sync, 86 types, 11 video, 86 view, 33, 85, 86 TRIGGER MENU button, 33 trigger position, readout, 25 trigger slope, indicator, 25 trigger source, 10, 26 TRIGGER VIEW button, 33, 85 triggering, basic concepts, 10
U utility, menu, 87 UTILITY button, 34
V vectors, 72 vertical menu, 89 position, 17 scale, 17 status, 88 video trigger, application example, 53 voltage cursors, 21, 71
volts/div coarse, 89 fine, 89 math, 78 VOLTS/DIV control, 30
W waveform See also applications acquiring data, 15 appearance, 28 compression, 75 displays, 28 expansion, 75 off, 90 saving and recalling, 81 taking measurements, 20 window time base, 32, 74 window zone, 26, 75
X XY, 72 application example, 64 display, 73
Y YT, 72
Index
120 TDS 200-Series Digital Oscilloscope User Manual