Fluke 189 (also 187) Undocumented Protocol

Two of Fluke's models of multimeters, the 187 and 189, include serial communications to transmit data to a computer. Partial information about the protocol used can be found here, among other places:

formatting link

However, the parts of the protocol documented by Fluke do not include any ability to fetch from the 189 the saved logs. I wrote to Fluke requesting additional information, but they could not provide it.

Thus, using the hint from the above web page that "QD 0" and "QD 2" were two of the commands of interest, I have spent some time revere engineering the protocol through extensive testing. The former provides information on the immediate reading, and may not be worth using unless you need information from the secondary display as well, as it is more work to decode than "QM", which provides similar information. The latter, however, is the only way of which I am aware to retrieve the logs.

The understanding presented below is incomplete. In particular, it does not address how to decode the information saying what mode the meter was set on. Also, there are a couple of other commands I have not investigated at all; since the documention mentions that the Fluke-supplied software (which I did not buy, as I do not have a Windows machine available anyway) can change settings like the logging interval on the meter, one or more of them presumably pertains to those operations. However, the understanding I now have is entirely adequate for my purposes (the downloading of logs), and I do not expect to work on it much more in the near future. I am therefore presenting it here in the hopes that it will save someone else the time of undertaking the same work, and that perhaps someone else will see fit to fill in the holes.

One might say "Shame on Fluke" for not providing it in the first place, and I could not argue, but sadly the fact that they were willing to provide any information whatsoever, however incomplete, makes them more helpful than some other companies I have tried to work with.

If you have any comments, reply here or email me (but note you will have to demunge the address to do the latter)

=========================================================================

Fluke 189 protocol undocumented commands. See here for the documented part and for the notes on the names of the other commands, which got me started:

formatting link

More information, maybe, if I've had time to write anything by the time you read this, at

formatting link

QD 0 response:

You transmit "QD 0\r". Fixed string of "0\rQD,", five bytes. All locations will be taken AFTER this.

Location Bytes Meaning 0 4 Time stamp in units of 0.1s 4 4 Primary display value. 8 1 # places to shift decimal point in primary value 9 1 Units. -1 for milli, 1 for kilo, etc. 10 4 Secondary reading. 14 1 # places to shift decimal point in secondary value 15 1 Units as above. 16 4 Tertiary reading. 20 4 Repeat of timestamp 24 4 Repeat of primary reading 28 1 Repeat of primary scale 29 1 Repeat of primary unknown 30 12 Function unknown. Presumably knob setting, etc.

All numerical values are little-endian.

Values with 0x70 in the high byte indicate that the value is not in use or that there is some kind of error. The low byte indicates what kind. 0x01 is for unused; other values indicate overflow, missing leads, etc. Otherwise they are 2s complement numbers.

Units indicate the prefix. -3 is nano, -2 is micro, -1 is milli, 1 is kilo, 2 is mega, etc.

The only use I have ever seen for the tertiary reading is the log # when in logging mode.

QD 2 response: If there ARE NO logs, response of "5\r". Otherwise, fixed string of "0\rQD,", five bytes. All locations will be taken AFTER this.

There is then an 18-byte header, which appears only once at the beginning. I know little about it, but presumably it includes information about the knob setting, etc.

Location Bytes Meaning 0 2 ? 2 4 Initial value 6 1 Decimal point location for initial value 7 1 Units, as in other places 8 2 Setting information, corresponding to QD0 bytes 30,31 10 2 ?, always 0x0258 that I can see 12 1 Mode information, corresponding to QD0 "mode" 13 5 ?, more unknown setting information as in QD0

The information about the setting is only partially understood.

Then there are 32 bytes of repeating information:

Location Bytes Meaning 0 4 Beginning timestamp for entry in units of 0.1s 4 1 Decimal point location 5 1 Units. -1 for milli, 1 for kilo, etc. 6 4 Minimum value for this time period 10 4 Maximum value for this time period 14 4 Actual value, sum of N 18 4 ?, always all 0 in my tests 22 4 Number of values which were summed 26 1 Status - 0x05 for normal, 0x85 for last entry, other values for range changes or other oddities 27 1 ? - always 0x01 in my tests 28 4 Ending timestamp for this entry

Generally, the readings are taken equally distributed over the time period, so the value will be an average. However, in cases of significant changes part way through, I have seen the meter use some heuristic to split it into separate entries sooner than the specified time period to capture this. In the case of a sharp voltage increase, I saw the first part (low value) have status 0x04, the short middle one with the fast increase had 0x08, and the final one had 0x05. The value displayed on the meter matched what I got by summing the actual values from the three and then dividing that by the sum of the number of readings from the three.

Reply to
Terran Melconian
Loading thread data ...

use

kind.

etc.

significant

to

first

the

and

three.

I have also tried to get info on this protocol. Since Fluke will not reveal their protocol, I purchase the next batch of DVM's from Amprobe. Although I have not tried to break down their protocol, or request it from them, they do offer a free download of their well written software package. I was able to do my captures in real time, and save them to csv files. From there I could do what I want. The only glitch is that all captures are @ 1 sample per second. But it is sometimes a quick solution, rather than draging out the $$$ heavy duty recorders! Oh, and Amprobe includes the IR to Serial cable with the meter, unlike Fluke!