Tektronix TDS460 error logs

Oct 06, 2010 2 Replies

Just finished repairing a TDS460, and I'm trying to clear the error logs. Sending ERR CLE or ERROR CLEAR via the GPIB, which works on the TDS5XX and up scopes does not seem to work on the TDS4XX scopes. Wondering if anyone knows how to clear the logs or what GPIB command is used? There's nothing in the service or programmers manual about this.



Thanks.


JW schrieb:

Ask here:

HTH

Reinhard

Thanks Reinhard, but already been there - I've been an active member there for a few years. Anyway, since I wasn't able to clear the error log via GPIB, I disconnected the battery for a minute while the power was off. On the first power up there was one power up failure on Dsp, and in the error logs all of the previous errors were gone. On subsequent power ups, all is well, except for the two errors that were generated from the loss of Vbat. (NVRAM and Pudiag failure.)

I tried to set the dip switch 3 off hoping like the TDS5XX scopes that it would skip these tests, but that didn't work either. Oh well, I guess I can live with just the two errors :)