Tek 475 trace/probe problem

Sep 11, 2013 24 Replies

Since you have 2 (or more) scopes, use the second one to look at the cal out on the noisy scope. This will rule out, or point to, a noisy cal generator. Be sure to get your grounds connected. I still think bad connector grounds are more likely on the noisy scope. Art

I've got the question on that group but so far more response here than there. I do have input from there and have eliminated a few things.

Mikek

OK. the $32 question. If you just short the input with no probe attached do you get this interference ?

What was the problem and solution? Art