There were users in some countries who did it. But it carries a high risk of infections and a huge liability risk for the hospital.

Regards, Joerg http://www.analogconsultants.com/

Too many pockets and crevices where blood could pool. You cannot irradiate this device becasue of electronics in there.

No, we as the device manufacturer said it.

?
Regards, Joerg http://www.analogconsultants.com/

what about autoclave? too hot? OTOH there's no money in making them reusable, until someone else does.

umop apisdn

You can't heat this stuff, else things could disintegrate immediately or worst case later. Parts coming off while in a coronary blood vessel would be a major situation.

I don't think anyone in their right mind will because that would mean major lawsuits waiting to happen. There are some devices for larger blood vessels that can be sterilized but I am not in favor of that and would not want to be involved in the design of those.

Regards, Joerg http://www.analogconsultants.com/

  • An, i presume autoclave is out of the question mainly due to hi temp.

Yes. Aside from semiconductors in this case, there are also bonding materials in there. An engineer is rather restricted in the choice of materials for such products because of biocompatibility and sterility requirements.

Regards, Joerg http://www.analogconsultants.com/