Tek 465 trace brightness variation during sweep

On my Tek 465 scope, I notice a peculiar variation of the intensity of the trace at fast sweep rates. The intensity decreases during the sweep for about 200 ns near the beginning of the trace and returns to normal for the remainder of the sweep. It is most noticeable using ALT sweeps.

Please see the pictures at

formatting link
formatting link
formatting link
formatting link

You can clearly see the intensity drop off for about 200 ns and come back up. For each picture, the inputs are the same; only the sweep rate was changed according to the name of the picture file. Its as if some of the sweep or trigger signal is leaking into the Z axis circuitry.

If you could suggest a cause (and remedy) for this I would be grateful. I do not have a service manual or schematic (I would like one), but I am generally good at troubleshooting and repairs. I just don't have any experience troubleshooting Tek scopes.

Thanks for any help you can provide with this.

-Al Schapira, KC2HRH, snipped-for-privacy@worldnet.att.net

Reply to
Al Schapira
Loading thread data ...

Something out of kilter in the bright-up/blanking/fly-back suppression circuitry ?

-- Diverse Devices, Southampton, England electronic hints and repair briefs , schematics/manuals list on

formatting link

Reply to
N Cook

Al Schapira wrote in news: snipped-for-privacy@worldnet.att.net:

Well,first thing to check is power supplies;voltage and ripple. Then,you will need a service manual with schematics;you need to scope the Z-axis waveform to see if it's flat at the time you get the intensity drop. The Z-axis amp and DC restorer is where I would be looking.

--
Jim Yanik
jyanik
at
kua.net
Reply to
Jim Yanik

Al,

Could I can't remember if a 465 does this or not, but HP scopes of that era would highlight the section of the trace that would be "magnified" by the delay when you had the delay mode switch set to normal and the delay ring of the timebase anywhere but off. Try adjusting the delay vernier to see of the highlighted section moves.

--
James T. White
Reply to
James T. White

I'd say James is right on the money, this looks like the scope is in "B intensify" mode.

Reply to
krikkit

"krikkit" wrote in news:3M8gf.206044$ir4.190226@edtnps90:

Then others here who own a 465 could duplicate the "problem".

--
Jim Yanik
jyanik
at
kua.net
Reply to
Jim Yanik

Thanks to all who replied.

. The A INTENSIFIED and B DELAYED controls work as expected. The scope was not in either mode when I took the original pictures.

. The problem seems to only appear in ALT sweep mode. When using CHOP to display the same waveforms, the intensity is uniform.

. Depressing the BEAM FINDER results in a pretty uniform intensity.

. When using A INTENSIFIED (by B) another similar anomaly occurs. After the intensified portion of the trace is supposed to return to normal intensity, it also drops off (dimmer than normal) for about 100 ns and then returns to normal. This, of course, happens, not just at the left edge, but anywhere the B DELAY is set to. From this I conclude that the Z Axis signal is overshooting/ringing, but only in the direction to dim the trace.

Since I don't have another scope fast enough to troubleshoot a 200 ns pulse (my other scope is a Tek 535A) I'll have to do it the hard way. Can anyone suggest which components might be involved?

Thanks again.

-Al

Al Schapira wrote:

...

Reply to
Al Schapira

Al Schapira wrote in news: snipped-for-privacy@worldnet.att.net:

First,I'd try to find out if other 465's do this.It may just be a quirk of the scope.It might just be the way your CRT grid bias is calibrated.

--
Jim Yanik
jyanik
at
kua.net
Reply to
Jim Yanik